BÜCHI Labortechnik AG
Operation | 6
Operation Manual Ultrasonic Package
23/34
u
Press the button
[Save]
.
ð
The output leveling mode is set.
6.2
Finding the optimal spray cone
Stall point
Run power
Low power
No droplet production.
High power
The liquid is being
ripped apart.
Optimal power
The liquid produces
droplets
6.2.1
Finding the optimal spray cone (Power-leveling mode)
Precondition:
R
Instrument is prepared for spraying.
u
Turn the navigation control clockwise until the nozzle starts producing droplets.
u
Turn the navigation control counter clockwise in 0.1 watt steps until the liquid
starts to spit from the tip of the nozzle and / or flooding the tip.
ð
The stall point is found.
u
Turn the navigation control 0.5 to 1.0 watt clockwise.
ð
The optimal spray cone is found.
6.2.2
Finding the optimal spray cone (Load-leveling mode)
Precondition:
R
Instrument is prepared for spraying.
u
Turn the navigation control clockwise until the nozzle starts producing droplets.
u
Turn the navigation control counter clockwise in 1 % steps until the liquid starts to
spit from the tip of the nozzle and / or flooding the tip.
ð
The stall point is found.
u
Turn the navigation control 2 - 5 % clockwise.
ð
The optimal spray cone is found.
6.3
Carrying out a spray drying process in open mode
6.3.1
Tasks before starting spray drying in open mode
Precondition:
R
u
Prepare 50 mL pure solvent.
u
Prepare the sample.
u
Place sample and solvent on top of the ultrasonic controller.
Содержание 11069891
Страница 1: ...Ultrasonic Package Operation Manual 11594107 A en...
Страница 33: ......