Safety Instructions
16
User Manual
Pos.
Safety equipment
Function
5
Instrument door with door
switches
Door switches connected to the safety circuitry ensure
that the beam shutter is closed immediately if the
instruments door is opened.
6
Primary radiation protection of
the X-ray tube
Housing made of 6 mm thick brass with a 3 mm wide
opening for limiting the beam emission
7
Beam shutter
Software-controlled beam shutter opened only for the time
of the measurement and only with closed instrument door
to minimize the time period of emitting X-rays.
8
Internal safety switches
Interlock circuit connected with the housing parts inside
the instrument. Opening the instrument for service
purposes will lead to an interruption of HV or even the
24 V power supply.
1 2 3 4 5 2 3 4 5
Fig. 1
Safety equipment on the front panel of the M1 ORA/MISTRAL
The key switch (1) for the M1 MISTRAL is located on the front panel and for the M1 ORA it can be
found at the back side close to the mains power supply connector. The safety equipment parts 2, 3,
and 4 are located on the front panel of the instrument (see Fig. 1). The safety equipment 6, 7 and 8
are located inside the instrument and can just be monitored and maintained by authorized service
staff.