S1 TITAN, T
RACER
5, and CTX User Manual
030.0112.04.0
13 of 67
Making a
filter
To make a custom filter –
Step
Action
1
Collect necessary tools:
Hole punch
Vacuum stick
0.050 inch hex screwdriver
Collimator assembly
2
Use the hole punch to cut the filter material.
3
Retrieve the filter piece from the hole punch.
4
Use the vacuum stick to pick up the filter piece
without touching it.
To operate the vacuum stick –
1. Press and hold the button.
2. Touch the tip to the filter piece.
3. Release the button.
5
Place the filter piece in the collimator. Release the
filter piece by pressing the vacuum stick button.
For a multilayer filter, the filter piece with the highest
atomic number must be placed in the collimator last
so it is closest to the X-ray source.
6
Screw the filter holder onto the collimator.
Your Complete Source for
Testing Equipment
www.BergEng.com
Berg Engineering & Sales Company, Inc.
1-847-577-3980