S1 TITAN, T
RACER
5, and CTX User Manual
030.0112.04.0
55 of 67
Toolbar displayed
Toolbar removed
1. ID match
If the sample matches grade IDs in the library, up to three matching IDs are displayed. If not, the
application name is displayed.
2. Assay
information
The following information is provided for quick reference –
Index number of the sample.
The date and time the assay was measured.
Length of the measurement in seconds.
If averaging, the index numbers of the selected assays.
3. Columns
Column Heading
Description
El
Element symbol.
Min and Max
Minimum and maximum percent allowed according to the grade
library. Note that some applications do not use a grade library and
Min
and
Max
are not displayed.
%, PPM, or Special Amount of the element in either percent, parts per million, or units
defined in EasyCal. See
EasyCal – a Comic Strip
, Bruker document
030.0101.
+/- [*n]
Standard deviation.
Sorting
To sort data in a column in descending or ascending order –
tap a column heading.
4. < and >
To view different assays ordered by index number –
use the back and forward options,
<
and
>
.
5. Info
To display the assay’s name, ID, and other information, if applicable –
tap
Info
. This information
cannot be edited.
1
2
3
4
4
5