S1 TITAN, T
RACER
5, and CTX User Manual
030.0112.04.0
46 of 67
8.1.
Display Settings
Description
How measurement results are displayed is specified on the
Display Settings
screen.
Description of
options
Option
Description
Display Units Measurement results can be displayed in –
PPM
– Parts per million.
%
– Percent.
Special
– An application-defined setting; or as defined in EasyCal. See
EasyCal – a Comic Strip
, Bruker document 030.0101.
Display Size
The font size on the Results screen.
Sort By
Results are sorted by element atomic number (
El No
) or alphabetically by
element name (
El Name
).
Enable LOD
Display
Results include analyzed elements with an assay less than the limits of
detection (LOD). The LOD is defined as n*STD, where n is a number from 1 - 5
and STD is standard deviation.
Number of
Decimals
The number of digits displayed to the right of the decimal point of a percent
value.
STD Display
Limit
The number of standard deviations on the Results screen. The error displayed
with an assay is n*STD. If the assay is less than n*STD and
Enable LOD Display
is checked, the assay is displayed as <LOD.