S1 TITAN, T
RACER
5, and CTX User Manual
030.0112.04.0
15 of 67
Changing
windows,
handheld
instruments
To change a window on the S1 TITAN or T
RACER
5 –
Step
Action
1
If necessary, unscrew and remove the nose plate using the
supplied screwdriver.
T
RACER
5
2
Peel off the old window.
Avoid any contact with the sensitive
detector.
S1 TITAN
3
Remove old adhesive with isopropyl alcohol.
T
RACER
5
4
Peel the white backing from the new window.
T
RACER
5
5
For the S1 TITAN, center alignment arrows over the middle of
the aperture.
S1 TITAN
6
Carefully apply the new window to the nose.
T
RACER
5
7
Peel off the clear front protective cover. Otherwise, light
element measurements will be inaccurate.
T
RACER
5
8
Replace the nose plate. Firmly secure the screws but do not
over-tighten.
Changing
To change a window on the CTX –
Your Complete Source for
Testing Equipment
www.BergEng.com
Berg Engineering & Sales Company, Inc.