7. SETTINGS Button – Spectrometer Mode
Description
If
Spectrometer Mode
was selected under
APPLICATION
, then under
SETTINGS
one of the following
screens is displayed, depending on instrument type.
S1 TITAN
T
RACER
5
Standard
settings
To select a standard spectrometer setting –
Step
Action
Result
1
From the
Ready to Test
screen with
application
Spectrometer Mode
selected, tap
SETTINGS
.
The
Spectrometer Settings
screen is displayed.
2
To select a defined setting
, tap the down
pointing triangle below
Reset
.
A dropdown list is displayed.
3
Tap a selection.
The setting name is displayed in the field and
the
kV
and
μ
A
fields reflect appropriate settings.
4
Tap the down pointing triangle to the
right of
Wheel
, scroll down if necessary,
and tap a filter name.
Note that S1 TITAN models 200 and 500
do not have
Wheel
displayed.
The filter name is displayed.
5
Define
System Settings
as necessary.
6
Tap
OK
.
Settings are saved and the
Ready to Test
screen
is displayed.