12. Quality Check
Description
This test verifies assays measured and recorded on the instrument are accurate.
Process
To verify that your instrument is
calibrated correctly
–
Step
Action
1
Set
APPLICATION
to the option applicable to the check sample supplied with the
instrument
and
Method
to
Automatic Calibration Selection
.
2
From the
Ready to Test
screen, measure the supplied check sample 10 times.
3
Average the 10 assays. See
Averaging
, page 58.
4
Compare the average measured value for each element against the Acceptance Limit
maximum and minimum on the Check Sample Report supplied with your instrument.
5
If all averaged measurements fall within MAX and MIN limits, continue to use the
instrument.
If averages fall outside MAX and MIN limits, with respect to standard deviation, check –
Measurement times
Sample surface
Instrument window
If the instrument does not perform as expected, contact Bruker at
, +1 (509) 783-9850.
Individual
reports
The following figure is an example only. When performing a quality check, use the Check Sample
Report included with your instrument. It contains values specific to it as determined during
testing at the factory.
Check Sample Report, example only