Technical Manual Version 002
BRUKER BIOSPIN
45 (55)
Authorised functions
WR - Write Record
4.8.18
Intel-Hex format is used to download the firmware on flash-eprom. The file to
transfer is generated with OHS51.EXE. Its file extension is «.HEX». This file is
composed by several records. Each record is composed as shown below :
:
L L A A A A T T D D D . . . . . . . . D D C C Cr Lf
Table 4.10.
Record format
Write
FIELD
LENGTH
SIGNIFICATION
: (3A)
1
Record start
L
2
Record length
A
4
Record address
T
2
Type (00: Data record, 01: EOF record)
D
LL
Data's
C
2
Checksum
Syntax :
EOT 0 0 0 0 STX
TR
Rec ETX BCC
Response :
CK if download initialized and the record processing issues.
NACK in all other cases
Description :
Allows to transfer records extract from a «.hex» file to the
BVT3X00.
Rules :
Rec value represents an intel-hex record.
The first character «:» (3A), Cr and Lf are not transmitted.