Technical Manual Version 002
BRUKER BIOSPIN
41 (55)
Authorised functions
P4 - Port 4 (for test only)
4.8.14
Port 4 allows to read the internal status of the BVT3200 and sets the gas flow rate
at power on.
It is composed as follows:
Table 4.9.
Port 4 definition
When a jumper is placed, it means that the valve is closed at power on.
Figure 4.2. Valve jumpers settings
Read
BIT
NAME
FUNCTION
0
unused
1
unused
2
unused
3
evap_on
1 = evaporator heater on
4
V1
jumper JP13
5
V2
jumper JP13
6
V3
jumper JP12
7
V4
jumper JP12
Syntax :
EOT 0 0 0 0
P4
ENQ
Response :
STX
P4
> ABCD ETX BCC
Description :
Allows direct read access to port 4.
Rules :
A and B are always «0». The third character represents the
state of port 3 bits 4 to 7. The fourth character represents the
state of the bits 0 to 3. All the characters are hexadecimal.