10 (55)
BRUKER BIOSPIN
Technical Manual Version 002
Table 1.1.
Default gas settings
Table 1.2.
Default gas flow versus jumper settings
Jumper
Valve
Pins to short
Factory setting (270l/h)
JP12
V4
pin 1(Gnd) - pin 2
Jumper placed
JP12
V3
pin 3(Gnd) - pin 4
No jumper
JP13
V2
pin 1(Gnd) - pin 2
Jumper placed
JP13
V1
pin 3(Gnd) - pin 4
Jumper placed
Combination
V1
V2
V3
V4
Flow rate (l/h)
0
0
0
0
0
0
1
0
0
0
1
135
2
0
0
1
0
270
3
0
0
1
1
400
4
0
1
0
0
535
5
0
1
0
1
670
6
0
1
1
0
800
7
0
1
1
1
935
8
1
0
0
0
1070
9
1
0
0
1
1200
10
1
0
1
0
1335
11
1
0
1
1
1470
12
1
1
0
0
1600
13
1
1
0
1
1735
14
1
1
1
0
1870
15
1
1
1
1
2000