26 (55)
BRUKER BIOSPIN
Technical Manual Version 002
Interface :
•
Serial link to Eurotherm 2416 controller :
9600 bauds,1 start bit, even parity, 1 stop bit and three wires link. Baudrate
can be changed by software with the «CO» command.
•
Serial link to host computer :
9600 bauds,1 start bit, even parity, 1 stop bit and three wires link.
Isolation :
Optocouplers 2500V isolation between digital interface and power section.
Power supply :
V = +5 Volt, I < 1 Ampere.
Commands and communication protocol
4.3
All commands for the Eurotherm controller cross over the interface. The micro-
controller looks at each received command and decides then for whom the
command is intended (either for the interface itself or the Eurotherm controller). A
command that is not an interface command is automatically transferred to the
Eurotherm controller. If the command is processed by the Eurotherm, the
controller answer is returned to the host computer via the interface.
Control characters
4.4
Six non printing ASCII characters are used to control the messages that are
exchanged between host computer and BVT3200.
Table 4.1.
Control characters
NAME
HEX
FUNCTION
STX
02
Start of text
ETX
03
End of text
EOT
04
End of transmission
ENQ
05
Enquiry
ACK
06
Acknowledge
NACK
15
Negative acknowledge
Содержание NMR BVT3200
Страница 16: ...16 55 BRUKER BIOSPIN Technical Manual Version 002...
Страница 24: ...24 55 BRUKER BIOSPIN Technical Manual Version 002...
Страница 50: ...50 55 BRUKER BIOSPIN Technical Manual Version 002 Figures...
Страница 52: ...52 55 BRUKER BIOSPIN Technical Manual Version 002 Tables...
Страница 55: ...Technical Manual Version 002 BRUKER BIOSPIN 55 55 End of Document...