40 (55)
BRUKER BIOSPIN
Technical Manual Version 002
P3 - Port 3 (for test only)
4.8.13
Port 3 allows to read the internal status of the BVT3200.
It is composed as follows:
Table 4.8.
Port 3 definition
Read
BIT
NAME
FUNCTION
0
heater on
1 = probe heater is ON
1
booster connected
1 = BVTB3500 present
2
evap conn
1 = evaporator device is connected
3
no gas
1 = missing gas flow
4
overheating
1 = overheating on probe heater
5
exch conn
1 = Exchanger connected
6
LN
2
refill
1 = LN
2
tank is almost empty
7
LN
2
empty
1 = LN
2
tank is empty
Syntax :
EOT 0 0 0 0
P3
ENQ
Response :
STX
P3
> ABCD ETX BCC
Description :
Allows direct read access to port 3.
Rules :
A and B are always «0». The third character represents the
state of port 3 bits 4 to 7. The fourth character represents the
state of the bits 0 to 3. All the characters are hexadecimal.