34 (55)
BRUKER BIOSPIN
Technical Manual Version 002
HP - Heater Power
4.8.7
Write
WARNING :
Space characters are not allowed.
Read
NB
: after power on the heater is OFF.
Syntax :
EOT 0 0 0 0 STX
HP
state
ETX BCC
Response :
ACK If state equals «0» or «1»
NACK In all other cases
Description :
This command allows to switch ON or OFF the gas flow
heating.
Rules :
State can be «0» or «1».
«1» switch the heater ON
«0» switch the heater OFF
All other values are ignored.
Syntax :
EOT 0 0 0 0
HP
ENQ
Response :
STX
HP
state ETX BCC
Description :
This command allows to read the heater's state.
Rules :
State can be «0» or «1».
«1» means that heater is ON
«0» means that heater is OFF