Technical Manual Version 002
BRUKER BIOSPIN
35 (55)
Authorised functions
IS - Interface Status
4.8.8
Read
Note :
Interface Status (IS) in the format (>ABCD)
Table 4.5.
Interface status
Syntax :
EOT 0 0 0 0
IS
ENQ
Response :
STX
IS
> ABCD ETX BCC
Description :
This command allows to read back the interface status.
Rules :
The status word is made of sixteen bits, each one represents a
particular function of the interface as detailed below.
The 16 bits are sent as four hexadecimal numbers preceded by
«>» to warn the computer that the data is hexadecimal.
Digits ABCD are ASCII characters representing a hexadecimal
digit (0-9, A-F).
DIGIT
BIT
SIGNAL NAME
FUNCTION
D1
0
heater on
1= heater is ON
D2
1
not used
always 0
D3
2
evap conn
1 = evaporator connected
D4
3
missing gas flow
1 = missing gas flow
C1
4
overheating
1 = heater overheating
C2
5
exch conn
1 = exchanger connected
C3
6
LN
2
refill
1 = refill LN
2
tank
C4
7
LN
2
empty
1 = LN
2
tank is empty
B1
8
evaporator status
1 = LN
2
heater is on
B2
9
not used
1 always
B3
10
booster connected
1 = BVTB3500 present
B4
11
reserved
0 always
A1
12
reserved
0 always
A2
13
reserved
0 always
A3
14
reserved
0 always
A4
15
reserved
0 always
Содержание NMR BVT3200
Страница 16: ...16 55 BRUKER BIOSPIN Technical Manual Version 002...
Страница 24: ...24 55 BRUKER BIOSPIN Technical Manual Version 002...
Страница 50: ...50 55 BRUKER BIOSPIN Technical Manual Version 002 Figures...
Страница 52: ...52 55 BRUKER BIOSPIN Technical Manual Version 002 Tables...
Страница 55: ...Technical Manual Version 002 BRUKER BIOSPIN 55 55 End of Document...