36 (55)
BRUKER BIOSPIN
Technical Manual Version 002
NH - Nitrogen Heater
4.8.9
Write
Read
NB
: Value is a DECIMAL code.
Syntax :
EOT 0 0 0 0 STX
NH
Value
ETX BCC
Response :
ACK
Description :
This command allows to control the LN
2
heater power
(Evaporator).
Rules :
The value from 0 up to 100%, is defined as a string up to 5
characters long. The string can begin with 1 to 5 spaces or «0».
After power on, the initial value is set to 0 (the nitrogen heater is
OFF).
Syntax :
EOT 0 0 0 0
NH
ENQ
Response :
STX
NH
Value ETX BCC
Description :
This command allows to read back LN
2
heater power.
Rules :
Value from 0 to 100%, is a string up to 5 characters long.
The string can begin with 1 to 5 spaces or «0».