Field Balancing
VIBROTEST 60
Page 114 of 168
C102 853.002
© VTEST60E July 2013 4.46ff
VIBROTEST 60 dialogue
VIBROTEST 60 steps
Activity at Rotor
Store the measurement with
»OK«.
4.
Store the Initial unba-
lance measurement
The measurement is stored with
»OK«.
With storage the trigger level is
determined in the VIBROTEST 60.
Even the smallest variations in the
relation between reference mark and
reference pickup will cause angle
errors.
The measurement can be repeated
with »START«
Stop the rotor
5.
End the Initial run
The Initial run can be repeated as
often as necessary with »SELECT«.
Test run
A
1
Select Test run with
»SELECT«.
1.
Select Test run in the
main menu
Open the test weight entry
menu with »EDIT«.
2.
Change the test
weight
The VIBROTEST 60 recommends a
test weight. The magnitude of the test
weight can be changed.
The test weight (TMA) is dis-played in
the measurement menu.