4.8 DECONTAMINATION CONT.
The following decontamination method is as
recommended in the ‘Code of Practise for the
Prevention of Infection in Clinical Laboratories
and Post-mortem Rooms’, ISBN 0 11 320464 7.
a. Bring the cryostat to room temperature.
b. Place 50-100ml of formalin BP in a flat dish
inside the chamber. Close the window.
c. Leave for at least 24 hours, preferably
48 hours.
d. Open the window and place a beaker
containing 10ml of ammonia SG.880 in the
chamber. Close the window.
e. Leave for one hour. The cryostat is
now decontaminated.
For further information regarding alternative
decontamination procedures please refer to
‘Safe Working and the Prevention of Infection
in Clinical Laboratories’, ISBN 0 11 885446 1.
4. Care & Maintenance
Cont.