Bright Instruments
46
4.6 SERVICING AND REPAIRS CONT.
If the cryostat or any part of it is returned
to the distributor or manufacturer, it must
be decontaminated (refer to section 4.8)
and cleaned thoroughly. Please remember
that the cleaning procedure should also
include removing the microtome and cleaning
underneath where it sits in the chamber.
A completed decontamination certificate
must be either sent in advance or attached
to the outside of the packaging of the
returned goods.
Work on the returned goods will not proceed
until the decontamination certificate has
been received (refer to details regarding
decontamination certificates on the page
marked Safety Information towards the
front of this instruction manual). Should no
decontamination certificate be received, or
the cryostat or any part of it be received in
a condition that Bright Instrument Co Ltd
consider to be a potential biological hazard,
the cryostat or part will be retuned,
un-repaired, at the expense of the customer.
4.6.1 UK
For customers in the UK, Bright Instrument Co
Ltd offer a comprehensive range of after sales
services that include extended warranties
and a full range of service contracts. For
further information or for any refrigeration,
electrical or mechanical problems contact
Bright Instrument Co Ltd direct providing the
following information:
• Cabinet Number (see ID plate on rear panel)
• Microtome Number (see ID plate on rear
panel)
• Date of installation
• Nature of fault
4.6.2 REST OF THE WORLD
Refrigeration problems are likely to be rare
and will normally be dealt with by a local
refrigeration specialist. For electrical and
mechanical problems contact your local
distributor of Bright products providing the
following information:
• Cabinet Number (see ID plate on rear panel)
• Microtome Number (see ID plate on rear
panel)
• Date of installation
• Nature of fault
4.7 SIMPLE SERVICING/REPAIR
PROCEDURES
All servicing must be carried out by a qualified
engineer. Any replacement parts must be
ordered from Bright Instrument Co Ltd - no
liability can be accepted if non-Bright supplied
parts are installed.
If there are any further problems contact
your local distributor or Bright Instrument
Co Ltd direct.
4.8 DECONTAMINATION
If decontamination is required carry out the
standard procedures as practised in your
laboratory. It is the responsibility of the
customer to use a decontamination procedure
appropriate to his/her work.
4. Care & Maintenance
Cont.