64 Setting Up the Vivo 40
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
Min and Max Inspiration Time (PSV only, optional)
CPAP (CPAP only, mandatory)
The Ramp Function (optional)
Setting range: Off, 0.3 sec [min] and
3 sec [max].
Navigate to the section “Setup” and
press “Setup” one more time to reach
the “More Settings” page.
Setting range: 4 to 20 cmH
2
O.
Navigate to the section “Setup”.
Ramp function: On, Off or Disabled
(where On and Off turns on and off
the ramp function in operating mode.
With the ramp function Disabled, it
cannot be activated in operating
mode).
Ramp time setting range: 10 to
60 min.
Start pressure setting range: 2 cmH
2
O to EPAP (PSV & PCV), 3 cmH
2
O to
CPAP (CPAP).
“Always start with ramp” setting range: On or Off or Disabled.
Navigate to the section “Others” and select “Ramp” to reach the “Ramp” page.
In home mode the ramp can be activated by pressing the ramp soft key for more
than 1 second.