93
Technical Specifications
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
High pressure
alarm
Self adjusting
Red LED, audible alarm
and a warning message
on the display.
Low volume
alarm
0.03 l to 2.0 l
resolution 0.05 l
Red LED, audible alarm
and a warning message
on the display. Accuracy:
0.05 l.
Low breath rate
alarm
4 BPM to 50 BPM
resolution 1 BPM
Yellow LED, audible
alarm and a warning mes-
sage on the display.
High breath rate
alarm
10 BPM to 60 BPM, Off
resolution 1 BPM
Yellow LED, audible
alarm and a warning mes-
sage on the display.
Low leakage
alarm
Self adjusting
Yellow LED, audible
alarm and a warning mes-
sage on the display.
High leakage
alarm
Self adjusting
Red LED, audible alarm
and a warning message
on the display.
Low Internal
Battery Warning
16.3 V
Yellow LED, audible
alarm and a warning mes-
sage on the display.
Low power
alarm
AC power: 65 ±15 V AC
Ext. DC 12 V: 10.0 ±0.5 V
Ext. DC 24 V: 20.0 ±0.5 V
Int. Batt.: 14 ±0.75 V
Red LED, audible alarm
and a warning message
on the display.
Internal function
failure alarms
Red or yellow LED, audi-
ble alarm and a warning
message on the display.
P
OWER
SUPPLIES
S
PECIFICATION
AC power supply
100 to 240 V AC, tolerance: +10%/-20%, 50
to 60 Hz, max 140 VA
A
LARM
S
PECIFICATION
I
NDICATION