62 Setting Up the Vivo 40
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
Breath Rate (PSV & PCV only, mandatory)
Inspiration Time (PCV only, mandatory)
Rise Time (PSV & PCV only, mandatory)
Inspiratory Trigger (in PSV mandatory, in PCV optional)
Setting range: 4 to 40 BPM.
Navigate to the section “Setup”.
Setting range: 0.3 to 5 seconds.
Navigate to the section “Setup” and
press “Setup” one more time to reach
the “More Settings” page.
Setting range: 1 to 9.
Navigate to the section “Setup” and
press “Setup” one more time to reach
the “More Settings” page.
In PSV mode, Rise Time is limited by
Min Inspiration Time.
Setting range: 1 to 9, Off (where 1 is
the most sensitive).
Navigate to the section “Setup” and
press “Setup” one more time to reach
the “More Settings” page.