83
Cleaning the Vivo 40 and Replacement of Accessories
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
10 Cleaning the Vivo 40 and Replacement of
Accessories
The patient-connected parts and the filter must be cleaned and replaced regu-
larly to ensure correct function of the Vivo 40. All replaced parts must be dis-
posed of in accordance with local environmental regulations regarding the
disposal of used equipment and waste.
10.1 Cleaning the Vivo 40
Main Unit
To avoid electrical shock, disconnect the ACpower supply to the Vivo 40
before cleaning. Do not immerse the Vivo 40 into any fluids.
• Always be careful when cleaning to ensure that you do not damage any
equipment.
• Fluid must not be allowed to enter into the Vivo 40.
• Do not sterilize the Vivo 40.
1
Switch off the Vivo 40 and dis-
connect the power supply.
2
Remove the patient circuit.