59
Setting Up the Vivo 40
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
7.1 Settings Applicable for the Different Modes
S
ETTING
R
ANGE
PCV PSV CPAP
EPAP
2 cmH
2
O to
20 cmH
2
O or IPAP/
IPAPmin -2 cmH
2
O.
IPAP 4 to 40 cmH
2
O
IPAPmin (Target
Volume active)
4 to 40 cmH
2
O or to
IPAPmax
CPAP
4 to 20 cmH
2
O
Breath rate
4 to 40 BPM (Breaths
Per Minute)
Inspiration time
0.3 to 5 sec
Min inspiration time Off, 0.3 to 3 sec
Max inspiration time
0.3 to 3 sec, Off
Rise time
1 to 9
Insp. trigger (PCV)
1 to 9, Off
Insp. trigger (PSV)
1 to 9
Exp. trigger
1 to 9
Target Volume
0.2 to 1.5 l
IPAPmax (Target
Volume active)
4 or IPAPmin
to 40 cmH
2
O
Ramp time (PSV,
PCV, CPAP)
10 to 60 min
Ramp start pressure
2 cmH
2
O to EPAP
Ramp start pressure
3 cmH
2
O to CPAP
HA 01 humidifier
1 to 9, Off
The ventilation modes and setting parameters are described in detail in the chapter
“Functions and Parameters of the Vivo 40” on page 27.