94 Technical Specifications
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
Internal battery
Capacity 3.8 Ah. NiMH (Nickel-Metal
Hydride). Operational time 3 hours, lifetime
3 years.
External battery
12/24 V DC, tolerance: +20%/-15% (10.5 to
15 V/20.4 to 30 V).
Max 120 W with Breas external battery.
E
NVIRONMENTAL
C
ONDITIONS
S
PECIFICATION
Operating temperature
range
41 to 100°F (5 to 38°C)
Storage and transport tem-
perature
-4 to +140°F (-20 to +60°C)
Ambient pressure range
600 to 1060 cmH
2
O, corresponding to
~4000 metres above sea level to ~375 metres
below sea level, at normal atmospheric pres-
sure.
Humidity
10% to 95%, non-condensing
O
PERATING
C
ONDITIONS
S
PECIFICATION
Recommended leakage
20 to 40 liter/min at 10 cmH
2
O
Minimum leakage
12 liter/min at 4 cmH
2
O
D
IMENSIONS
S
PECIFICATIONS
W
×
H
×
D
7.48
×
9.57
×
8.78 inch (with rear lid)
Weight
8.8 lbs (with rear lid)
Air outlet
22 mm male conical standard connector
P
OWER
SUPPLIES
S
PECIFICATION