76 Alarms
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
8.3 Technical Alarm
Low Internal Battery Warning
The low battery alarm can be triggered prematurely by a sudden increase
in flow, such as an excessive leakage, a large breath or a mask off event. If
this should happen, restart the Vivo 40 on internal battery. If the low bat-
tery alarm persists, the internal battery needs to be charged.
I
TEM
D
ESCRIPTION
Definition
A low internal battery warning will be given when
the internal battery is the last power source, and it
falls below the voltage warning limit. Under nor-
mal circumstances, the low battery alarm will be
activated approximately 15 minutes before shut-
down.
Possible cause
A discharged internal battery.
Warning limit
Internal battery: 16.3 V
Ventilator action
The Vivo 40 will continue to give breaths with the
same settings.
Indication
The warning is given audibly with a tone and visi-
bly by the yellow alarm LED and a display mes-
sage.