91
Technical Specifications
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
CPAP
4 to 20 cmH
2
O.
Tolerance: ±2% of the maxi-
mum value and ±10% of the
set value.
0.5 below 10 cmH
2
O
1.0 above 10 cmH
2
O
Breath rate
4 to 40 breaths per minute
(BPM), tolerance: ±10% of
set value.
1 BPM
Inspiration time
0.3 to 5 sec, tolerance: ±10%
of set value.
0.1 sec
Min inspiration
time
Off, 0.3 to 3 sec
0.1 sec
Max inspiration
time
0.3 to 3 sec, Off
0.1 sec
Rise time
1 to 9
1
Inspiratory trig-
ger effort level
1 to 9, Off.
1
Expiratory trig-
ger effort level
1 to 9, where 1 is the lowest
effort and 9 is the highest
effort setting.
1
Ramp function
On, Off, Disabled
HA 01 humidi-
fier
Settings: Off, 1 to 9, corre-
sponds to 10 to 30 mgH
2
O/l,
<100% RH. Heat-up time
from 73°F (23°C): less than
1 hour. Max gas temperature
at patient port: 109°F (43°C).
1
Audible alarm
level
1 to 9, where 1 is the lowest
and 9 is the highest volume
setting.
1
Maximum flow
> 200 liter/min
S
ETTING
/V
ALUE
R
ANGE
/P
ERFORMANCE
R
ESOLUTION