Unit Diagnosis / Calibration
2
2 - 4
Infusomat® Space 6.0
EN
Error Codes of the Control Microprocessor
*
* Note: If alarm 2111 occurs during infusion, the device history logs may not be re-
trievable.
Error Code
Definition
Possible Cause
Fault Clearance
1
2100
Timebase too fast
Quartz of the processor PCB
Exchange processor PCB
2
2101
Timebase too slow
Quartz of the processor PCB
Exchange processor PCB
3
2102
Timebase failure
Quartz of the processor PCB
Exchange processor PCB
4
2103
Keyboard High
Keyboard defective
Exchange operating unit
5
2104
ON/OFF key pressed too long
Keyboard defective
Exchange operating unit
6
2105
No keydecode
Keyboard defective
Exchange operating unit
7
2106
ROM Romtest defective
Software
Update unit software
8
2107
ROM program defective
9
2108
FuP has not changed active reset value
10 2109
MPU_Test failed
11 2110
RAM_Test failed
12 2111*
KUP active reset
Pump was not switched off cor-
rectly
Restart device
13 2112 2114
Internal Error
14 2115
Drive too fast
Motor drive
Detection of direction of rotation
Exchange processor PCB
15 2116
Drive too slow
16 2117 ... 2118 Internal Error
17 2119
LCD backlight on defective
LC display defective
Exchange operating unit
18 2120
LCD backlight off defective
19 2121
Red LED on defective
20 2122
Red LED off defective
21 2123
Key pressed too long (without ON/OFF
key)
Keyboard defective
Exchange operating unit
22 2124 ... 2127 Internal Error
23 2128
Drive motion incorrectly forward
Motor drive
Detection of direction of rotation
Replace processor PCB
24 2129
Drive motion incorrectly backward
25 2130 ... 2159 Internal Error
26 2160 ... 2162 PCA Error
Error in relation to PCA therapy
together with special settings
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
27 2170 ... 2173 DLOCK Error
Error in relation to Datalock to-
gether with special settings
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaSeD and inform B. Braun
Table 2 - 3 Error Codes of the Control Microprocessor (Part 1 of 3)
Содержание 0871 3050
Страница 1: ...Infusomat Space Service Manual Version 6 0 English 0 12 0 RX only...
Страница 12: ...Contact Persons 0 0 12 Infusomat Space 6 0 EN For your notes...
Страница 20: ...System Overview 1 1 8 Infusomat Space 6 0 EN Fig 1 7 Block Diagram Infusomat Space...
Страница 30: ...System Overview 1 1 18 Infusomat Space 6 0 EN For your notes...
Страница 108: ...Servicing the Unit 4 4 4 Infusomat Space 6 0 EN For your notes...
Страница 130: ...Spare Parts List 8 8 6 Infusomat Space 6 0 EN For your notes...
Страница 135: ......