Disassembly / Assembly
3
3 - 58
Infusomat® Space 6.0
EN
Air inline sensor
(see TSC for limit values)
Infusomat® Space Line, standard type
❒
Water value
❒
Difference of temperature sensors
❒
Air alarm
❒
Air value
Note:
If the water or air value is not
reached immediately, these values have to
be read again after 120 seconds.
Switch on unit without power supply
❒
Self-test
❒
Magnetic function of the battery com-
partment cover
❒
Charging level of the battery module
Remove battery module
❒
Alarm for at least 3 minutes.
Visual Inspection
Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection
(Part 3 of 3)
Date:
Name:
Signature: