Functional Inspection
Infusomat® Space
Mechanical Inspection
1. Fit the unit to be tested on top of another Space device and
check the proper functioning of the lock.
2. Fit the unit to be tested under another Space device and
check the proper functioning of the lock.
3. Check the battery compartment cover lock for proper opera-
tion.
Functional Check
1. Determine age of battery
Open battery compartment cover and remove battery.
Read batch number on the battery nameplate and determine
age of the battery:
a) Please see
for NiMH batteries.
b) Please see
for Li-ion batteries.
Note
If the battery is more than 8 years old, recommend changing the
battery or conducting a battery check.
Fig.: 6 - 2 NiMH batteries
Fig.: 6 - 3 Li-ion batteries
Letter: Year of manufacture
according to table left
Two digits: Calendar week
Example:
Manufactured in week 42/2008
D 2003
E 2004
F 2005
G 2006
H 2007
I 2008
...
N 2013
First two digits: Calendar week
Second two digits: Year of manufacture
Example:
Manufactured in week 41/2011
Содержание 0871 3050
Страница 1: ...Infusomat Space Service Manual Version 6 0 English 0 12 0 RX only...
Страница 12: ...Contact Persons 0 0 12 Infusomat Space 6 0 EN For your notes...
Страница 20: ...System Overview 1 1 8 Infusomat Space 6 0 EN Fig 1 7 Block Diagram Infusomat Space...
Страница 30: ...System Overview 1 1 18 Infusomat Space 6 0 EN For your notes...
Страница 108: ...Servicing the Unit 4 4 4 Infusomat Space 6 0 EN For your notes...
Страница 130: ...Spare Parts List 8 8 6 Infusomat Space 6 0 EN For your notes...
Страница 135: ......