Disassembly / Assembly
3
Infusomat® Space 6.0
3 - 35
EN
Repairing Metal Front Sheet, Outsert ISPS, Version II
Designation
Order No.
Front flap sealing strip. . . . . . . . . . . . . . . . . . . . . . . . . . On request
1. Loosen the six screws.
The screws can be reused, but use new sealing disks.
Note
The keyboard screws (
) must not be loosened.
A special procedure is required to install the keyboard, so that a
uniform pressure point is guaranteed for all keys.
2. Open the rear wall of the operating unit carefully (silicone
seal). Pay attention to the chassis earth of the keyboard cir-
cuit board.
3. Ensure that the flexible strip is not damaged.
4. If the silicone seal is damaged, remove the loose residues.
5. Bend the earthing wire upward.
6. Fold up the bonded protection at the sealing strip for the
front hood at both ends approx. 1 cm.
7. First bond the sealing strip in the corners.
8. Always roll and bond the bonded protection only approx.
5 cm.
Fig.: 3 - 34
Legend of fig. 3 - 34:
Item Designation
1 Screw
Fig.: 3 - 35