150kHz to 80MHz
d=1.2 P
4
Recommended separation distances between portable and mobile RF communications
equipment and the EndoSequence II.
The EndoSequence II is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the EndoSequence II can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the EndoSequence II as recommended below,
according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Rated maximum output
power of transmitter
W
0.01
800MHz to 2.5GHz
d=2.3 P
80MHz to 800MHz
d=1.2 P
0.12
0.23
0.12
0.1
0.38
0.73
0.38
1
1.2
2.3
1.2
10
3.8
7.3
3.8
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
100
12
23
12
Separation distance according to frequency of transmitter
m