background image

NOTICE

Liquid crystal panel

ON/OFF key

PROGRAM key

DOWN key

POWER key

UP key

SELECT key

<Operation Panel>

16

(4) Preparatory Operations

1. Hold down the POWER key for more than one 

second to turn on the power.

2. Press the PROGRAM key until program No. suitable 

for file to be used is displayed.

3. When changing the set value of Speed, Torque, 

Gear Ratio or Auto Reverse, press the SELECT key, 
select the setting item and press the UP/DOWN key 
to change.

(5) Operation

If you press the ON/OFF key briefly, the motor 
handpiece starts. If you re-press the key, it stops. 
(Alternate operation)
If you hold down the ON/OFF key for more than one 
second, the motor handpiece starts while the key is 
pressed. If you release the key, it stops. (Intermittent 
operation)

    

Auto Reverse Function

Auto reverse setting can be changed by pressing the SELECT key for one second or longer 
while the motor handpiece stops. As the auto reverse mark flashes while the SELECT key is 
kept pressed, press the UP/DOWN key to adjust it.

: Auto Reverse

: Auto Stop

: Auto Reverse OFF (flashes while the SELECT key is kept pressed, and goes out 
when it is released) 

The alarm will sound when it becomes a value of half of a torque limit set up during motor 
handpiece rotation, and the alarm will change near the torque limit value. (Load alarm) When 
it continues applying a load and it exceeds the torque limit value you can select mode from 
the following three modes.

The GEAR RATIO key is Fast-Forwarded, when it is 
continuously held down.

Содержание EndoSequence II

Страница 1: ...OPERATION MANUAL OM ER436E Cordless Handpiece...

Страница 2: ...l Do not use this for implants other than endodontic treatment or other dental treatment Guidance and manufacturer s declaration electromagnetic emissions The EndoSequence II is intended for use in th...

Страница 3: ...les 70 Ut 30 dip in Ut for 25 cycles 5 Ut 95 dip in Ut for 5 secs 3 A m Compliance level 6kV contact 8kV air 2kV for power supply lines 1kV line to line 2kV lines to earth 5 Ut 95 dip in Ut for 0 5 cy...

Страница 4: ...ld be observed to verity normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the EndoSequence II b Over the frequency range 150k...

Страница 5: ...ers rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the...

Страница 6: ...product properly These indicators are to allow you to use the product safely and prevent danger and harm to you and others These are classified by degree of danger damage and seriousness All indicator...

Страница 7: ...es Do not drop the motor handpiece or battery charger Place the battery charger on a flat and stable surface Should the battery fluid leak and get into your eyes immediately wash eyes thoroughly with...

Страница 8: ...rce Before changing the head or file turn off the power of the motor handpiece Changing with the power on may cause unintended rotation by accidental activation of the ON OFF switch When inserting the...

Страница 9: ...and operating time become significantly shorter since the charging current differs The motor handpiece consumes electricity very slightly even when the power is off In addition fully charged rechargea...

Страница 10: ...tion The built in feedback circuit which keeps rotation at a constant speed even when the load on the motor handpiece changes Non contact charger prevents improper charging due to deterioration of met...

Страница 11: ...Pressing this switch when the power is on rotates the motor handpiece and de pressing the switch again stops the handpiece Normal rotation operation Pressing and holding the switch for approx one seco...

Страница 12: ...rque suitable for contra angle which you use For gear ratio 4 1 10 1 16 1 or 20 1 can be set LCD Panel AUTO REVERSE The current auto reverse mode mark is displayed AUTO REVERSE If the load is removed...

Страница 13: ...are drained or in a remarkably low voltage Charge the batteries ALARM Symbol The present alarm symbol is displayed Alarm ON Alarm OFF No display The symbol indicating the remaining amount of the batt...

Страница 14: ...this time check that the power lamp lights up Fig 5 4 Insert the motor handpiece into the Battery Charger Charging starts with the charge mark flashing on LCD screen 5 When the alarm sounds and is dis...

Страница 15: ...s charging cannot be properly completed and the battery compartment may feel hot We recommend that you charge batteries for longer periods if possible The power of the motor handpiece should not be tu...

Страница 16: ...it engages with the latch mechanism Push it inward to click Fig 7 File Removal depress the push key and pull out the file Fig 8 When attaching and detaching the file turn off the power beforehand Aft...

Страница 17: ...or handpiece starts while the key is pressed If you release the key it stops Intermittent operation Auto Reverse Function Auto reverse setting can be changed by pressing the SELECT key for one second...

Страница 18: ...ad higher than the set torque limit value AUTO REVERSE The handpiece starts in forward rotation When a load higher than the torque limit is applied the file will rotate in reverse When the load is rem...

Страница 19: ...Calibration This function is to decrease fluctuation in the rotation speed of the motor handpiece and the difference in torque by the contra angle head 1 Lubricate the contra angle head with BRASSELER...

Страница 20: ...5 Display on the liquid crystal panel LCD returns to the original state if no operation is performed for a while This function does not work unless remaining battery capacity is sufficient Perform ca...

Страница 21: ...and it may flows backward 2 Cleaning Motor Handpiece When the motor handpiece becomes dirty wipe it off with a cotton cloth moistened with rubbing alcohol 3 Cleaning Charger When the charger becomes d...

Страница 22: ...in at 121 C 250 F or 15 min at 132 C 270 F 5 Keep the handpiece in the autoclave pouch to keep it clean until you use it Sterilization at 121 C for more than 15 minutes is recommended by EN ISO 17665...

Страница 23: ...sulting from you changing the batteries yourself and not following the CAUTIONS ON CHANGING BATTERIES 11 Changing Batteries Do not open any part other than the battery cover Use only batteries as spec...

Страница 24: ...e old batteries 6 Insert new batteries according to the polarity marking in the battery box A mix up between the plus and minus sides will not allow operation of the handpiece 7 Attach the battery cov...

Страница 25: ...on of the circuit may be suspected Contact Brasseler USA Replace the motor handpiece or contra angle head Malfunction of circuit The motor handpiece is locked at the time of the auto reverse mode Malf...