LAST SETTING
When the power is turned OFF the last setting used is memorized. When the Power is turned
back ON, the most recent settings will appear.
Auto Power Off
If the handpiece is not operated or approximately 10 minutes passes during rotation with
no-load, the power is automatically turned OFF to save energy and prevent improper
operation. However, if load is applied during rotation of the motor handpiece, the power is not
turned OFF even during no operation.
NOTICE
18
(1) Program
You can change any preset values and have them memorized to your desired settings
(rotation speed, torque limit value, gear ratio and auto reverse mode)
1. Press the PROGRAM key until it turns to the program number which you want to have
memorized.
2. Adjust the rotation speed, torque limit value, gear ratio and auto reverse mode by
UP/DOWN key according to your needs.
3. Hold down the PROGRAM key for more than one second. When the alarm sounds the
settings have been memorized.
(2) Calibration
This function is to decrease fluctuation in the rotation speed of the motor handpiece and the
difference in torque by the contra angle head.
1. Lubricate the contra angle head with BRASSELER USA SPRAY).
(Refer to the “7 (1) Lubricating Contra Angle Head.”)
2. Turn on the power.
3. Hold down UP/DOWN key simultaneously for more than one second.
4. The LCD panel displays “ ” with the alarm sound.
5. Attach the contra angle head to the motor handpiece and press the ON/OFF key.
(Remove the file or TEST bur.)
6. If the motor handpiece begins to rotate, leave it as it is until it stops.
7. This process ends, if the rotation stops and the LCD panel display returns to its original state.
6. Convenient Function
(6) Completion of Medical Treatment
When the treatment is completed, hold down the POWER key for more than one second to
turn off the power and return the motor handpiece to the handpiece stand
• The program cannot be memorized while the motor handpiece is in motion.
• The program is not memorized unless the PROGRAM key is held down for more than
one second. If the program number is changed by the PROGRAM key, the initially
memorized preset values remain. (Cancel function)