11
• PROGRAM Key
A program can be selected.
The present program number is displayed. There are 5 programs available, numbered
from 1 to 5.
Up to five programs can be set, and program settings can be memorized by pressing this
key for one second or longer.
The following parameters can be set: Speed, Torque, Gear Ratio and Auto Reverse setting.
• UP/DOWN Key
Use this key when adjusting the set value for each parameter.
Alarm sounds if the set value exceeds the upper limit or lower limit.
Unit to be set is min
-1
for speed and N
·
cm for torque. Set torque suitable for contra angle
which you use. For gear ratio, 4:1, 10:1, 16:1 or 20:1 can be set.
<LCD Panel>
• AUTO REVERSE
The current auto reverse mode mark is displayed.
AUTO REVERSE:
If the load is removed after auto reverse rotation, it returns to the
normal rotation again.
AUTO STOP:
If the load is removed after auto reverse rotation, it stops.
AUTO REVERSE OFF:
Auto reverse rotation is not activated.
• If the speed or torque set value has reached the upper limit or lower limit when the
gear ratio has been changed, the alarm sounds.
• The indication of “ ” for torque setting is the upper limit value of torque. If you
attempt to set torque which exceeds this value, the alarm sounds.
• The speed can be set at any time if it is normally indicated.