*Speed = Rotation speed
of the file
10
4. Parts and its function
Program No.
Speed*
Torque
Auto Reverse mode
Buttery Symbol
Alarm Symbol
Gear ratio
(
Gear ratio setting screen
)
(
Torque setting screen
)
<LCD Panel>
LCD Panel
ON/OFF Key
PROGRAM Key
SELECT Key
DOWN Key
UP Key
Power Key
<Operation Panel Details>
<Operation Panel Details>
• POWER Key
- Holding down the POWER key for more than one second turns on the power and the LCD
panel lights.
- Holding down the POWER key for more than one second while the power is on turns off
the power and the LCD display turns off.
• ON/OFF Key
Pressing this switch, when the power is on, rotates the motor handpiece, and de-pressing
the switch again stops the handpiece. (Normal rotation operation) Pressing and holding the
switch for approx. one second or longer rotates the motor handpiece and releasing the
switch stops the handpiece. (Temporary rotation operation)
• SELECT Key
Press the SELECT key to change speed, torque or gear ratio setting. Select the parameter
for which you want to adjust the set value.
The following parameters can be changed in the following order: speed (SP)
torque (TQ)
gear ratio (GR). However, gear ratio cannot be set during rotation of handpiece.
Pressing and holding this SELECT Key for 1 second or longer, when the motor handpiece is
stopped can change the Auto reverse mode. (See Auto Reverse)