CAUTION
Motor Cap
Motor Handpiece
21
Sterilize the contra angle head only.
For the sterilization method, we recommend autoclaving. Autoclave sterilization required the
first time you use and after each patient as noted below.
Autoclaving
1. Brush the dirt off the surface of the contra angle head, and wipe it off with a cotton cloth
moistened with isopropyl alcohol. Do not use a metal brush.
2. Lubricate the head with the SPRAY.
3. Insert the head into an autoclave pouch and seal it.
4. Autoclavable up to a max. 135°C (275°F).
ex.) Autoclave for 20 min. at 121°C (250°F), or 15 min. at 132°C (270°F).
5. Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* Sterilization at 121°C for more than 15 minutes is recommended by EN ISO 17665-1
8. Sterilization
• Do not autoclave the motor handpiece.
• Since the lowermost tray inside the chamber of the autoclave is close to a heater and
the temperature in that place may locally exceed 135°C (275°F), place the head on
the central or upper tray.
• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
When the contra angle head is removed from the
motor handpiece for battery charging,
lubrication, or sterilization, mount the motor cap
onto the motor handpiece to prevent debris from
entering.
9. Motor Cap
Fig. 12