Technical Manual
3-18
Range/Alternatives
Explanation and
<default value>
result of alternatives.
1:No Motion Delay
Range:
0 to 10.0
Unit: s
<1.0>
Delay in seconds from detection of stable weight
until the Motion status goes off.
1:Motion Check
Off
On
<Off>
Off: Only zero adjustment is inhibited during motion.
On: Inhibits zero adjustment, taring, and printing during
motion.
1:Min. Weight Print
Range:
0 to 999999
Unit:
Measurem. unit
<0.0>
Defines the smallest allowed weight value to be printed.
0 gives no restrictions.
1:Overload Check
Off
Unipolar
Bipolar
<Off>
Check of overload is performed according to this set-up.
Off: No check is performed. Weight will be reported
up to the limits of the AD converter.
Unipolar: Overload status will be set if ‘Overload limit’ is
exceeded. Underload status will be set if the gross
weight is below minus (-)9 * Resolution.
Bipolar: Overload status will be set if Overload limit
is exceeded.
Underload status will be set if the gross weight is below
minus (-) Overload limit.
1:Overload Limit
Range:
0.5 to 999999
Unit:
Measurem. unit
<Ca
9 * Resolution>
Should be set to the max gross weight that the weight
indicator is allowed to report. This parameter is always
set to default value when changing Resolution or
Capacity. If ‘Bipolar’ overload check is selected the
weight will be reported up to the Overload limit and down
to the minus (-) Overload limit.
Note: this parameter is only shown if parameter
‘1: Overload Check’ is set to ‘Unipolar’ or ‘Bipolar’.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Weighing Instrument Program version 1 12 0 0 Technical Manual PM DT HE types ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 41: ...G4 Multi Channel Weighing Instrument 3 5 ...
Страница 154: ...Technical Manual 6 48 ...
Страница 180: ......
Страница 190: ...Technical Manual 9 10 ...
Страница 191: ...G4 Multi Channel Weighing Instrument Appendix 1 ...
Страница 192: ...Technical Manual ...
Страница 193: ......