DC SUPPLY
The output of the external DC source must be rated 24 V
, ±15%
including fluctuations, min. 40 W. The DC source must provide Double
Insulation between Mains parts and 24 V SELV or SELV-E Circuit,
and a limited-energy circuit (maximum available current of 8 A).
For the US market this energy limit can be achieved with an
ANSI/UL248-14 fuse rated 5A.
For other markets an IEC 60127 T type fuse rated 4A may also be used.
24 V
In
Terminals 1, 2 and 3. The G4 instrument should be powered by 24
V
, connected according to the diagram below. To achieve
functional grounding, terminal 3 should be connected to ground.
See Technical data for input voltage ratings.
24 V
Out
Terminals 4 and 5 can be used to supply max. 100 mA to digital
inputs and outputs. Connection should be according to the diagram
below.
See technical data for ratings.
1
DC SUPPLY
2
0 V
3
Shield
4
5
0 V
6
Shield
0 V
Ground
0 V
+24 V In
+24 VDC In
+24 VDC Out
+24 V Out
!