G4 Multi Channel Weighing Instrument
9-9
Error
code
Explanation
192
Too high transducer signal in calibration point 2
The mV/V signal in calibration point two is too high (often due to a previous,
strange data sheet calibration).
193
Too high transducer signal in calibration point 2
The mV/V signal in calibration point 2 is too high, due to strange data sheet
calibration. The conversion factor, rated load etc. does not correspond to
each other.
194
Illegal calibration direction
All weights with corresponding mV/V values must be increasing for
increasing calibration point number.
195
Illegal I/O module configuration
The I/O-module set-up (Hardware Configuration menu) is not consistent
with the allowed number of modules of the different types or the allowed
combinations of modules.
196
Illegal scale number configuration
Each scale number must only be used once. I.e. each measuring input must
have a unique scale number. See menu ‘Hardware Configuration’.