![Blackrock Microsystems NeuroPort Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/blackrock-microsystems/neuroport/neuroport_instructions-for-use-manual_2767812034.webp)
Revision 1.00 / LB-1070 – NeuroPort Biopotential Signal Processing System IFU
© 2021 Blackrock Microsystems, LLC
34
Analog Output: Triggering
Sine and custom waveforms can be triggered by different conditions listed below.
Instant
: Waveform triggers when OK or Apply are clicked.
Digital Input Rising Edge
: Waveform triggers when a user defined digital input
pin transitions from low to high. A digital input bit must be selected.
Digital Input Falling Edge
: Waveform triggers when a user defined digital input
pin transitions from high to low. A digital input bit must be selected.
Spike Unit
: Waveform triggers when a sorted unit is detected on a defined
channel.
NeuroMotive ROI
: Triggers when a subject enters or exits a selected region of
interest.
Extension
: Waveform triggers based on custom extension code loaded onto
hardware.
Digital Input
Digital input properties are accessed from the left side bar of the Hardware
Configuration Window and allows the user to set the properties of the digital input
ports. Digital inputs can synchronize external hardware and pass digital data into
neural recordings. Digital Inputs may take on the functions listed below.
16-bit on word strobe
: Reads 16 bits on the rising edge of the strobe pin.
Figure 21–Analog Output: Custom