Trigger
57
State
Triggers when the signal enters or leaves a voltage level and stays there for a specified time. This is useful for detecting
when a signal gets stuck at a particular level.
Figure 6.22
Dropout Trigger State
To set a state dropout trigger:
1. Press the Setup key to enter the
Trigger
menu.
2. Press the
Type
softkey, then use the Universal Knob to select
Dropout
.
3. Press the
Source
softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source
4. Press the
Slope
softkey to select rising or falling edge.
5. Press the
OverTime Type
softkey to select
State
.
6. Press the
Time softkey
, then use the Universal Knob to set the desired value.
Note:
Coupling and noise reject can be set in interval trigger, see the sections
Trigger Coupling
and
Noise Rejection
for
details.