Trigger
43
6.6 Noise Rejection
Noise Rejection
adds additional hysteresis to the trigger circuitry. By increasing the trigger hysteresis band, the possibility
of triggering on noise is decreased, however the trigger sensitivity is also decreased. A larger signal is required to trigger
the oscilloscope when
Noise Rejection
is enabled.
To enable
Noise Rejection
:
1. Press the Setup key to enter the
Trigger
menu.
2. Press the
Noise Reject
softkey to toggle between On and Off.
Disabled
Enabled
Figure 6.6
Noise Reject
If the signal being probed is noisy, set up the oscilloscope to reduce the noise in the trigger path and on the displayed
waveform. First, stabilize the displayed waveform by removing the noise from the trigger path. Second, reduce the noise
on the displayed waveform.
• Obtain a stable display.
• Remove the noise from the rtigger path setting Trigger coupling to LF Reject,
HF Reject
, or enabling
Noise Reject
.
• Set the
mode to
Average
to reduce noise.
6.7 Trigger Types
The 2194 provides the following trigger types :
•
•
•
•
•
•
•
•
•
• Serial Trigger
Figure 6.7
Trigger Type