6.7.6 Interval Trigger
Trigger when the times difference between the neighboring rising or falling edges meets the time limit
(< =, > =, [ - - . - - ], - - ][ - - ).
Figure 6.19
Interval Trigger
To set an interval trigger:
1. Press the Setup key to enter the
Trigger
menu.
2. Press the
Type
softkey, then use the Universal Knob to select
Interval
.
3. Press the
Source
softkey, then use the Universal Knob to select CH1 or CH2 as the trigger source
4. Press the
Slope
softkey to select rising or falling edge.
5. Press the
Limit Range
softkey, then use the Universal Knob to set the desired condition.
•
< = (less than a time value)
: Triggers when the positive or negative pulse time of the input signal is lower
than the specified time value.
•
> = (greater than a time value)
: Triggers when the positive or negative pulse time of the input signal is
greater than the specified time value.
•
[ - - . - - ] (within a range of time value)
: Triggers when the positive or negative pulse time of the input
signal is greater than the specified lower limit of time and lower than the specified upper limit of time value.
•
- - ] [ - - (outside a range of time value)
: Triggers when the positive or negative pulse time of the input signal
is greater than the specified upper limit of time and lower than the specified lower limit of time value.
6. Press the
Time Setting
softkey
(< = , > = , [ - - . - - ] , - - ] [ - -
), turn the Universal Knob to select the