Measure
100
13.
Overshoot
: Overshoot is distortion that follows a major edge transition expressed as a percentage of Amplitude. ROV
means rising edge overshoot and FOV means falling edge overshoot.
Figure 11.2
Overshoot
𝑅𝑖𝑠𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑜𝑣𝑒𝑟𝑠ℎ𝑜𝑜𝑡 =
𝑙𝑜𝑐𝑎𝑙 𝑀 𝑎𝑥𝑖𝑚𝑢𝑚 − 𝐷 𝑇 𝑜𝑝
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝑋 100
𝐹 𝑎𝑙𝑙𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑜𝑣𝑒𝑟𝑠ℎ𝑜𝑜𝑡 =
𝐵𝑎𝑠𝑒 − 𝐷 𝑙𝑜𝑐𝑎𝑙 𝑀 𝑖𝑛𝑖𝑚𝑢𝑚
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝑋 100
14.
Preshoot
: Preshoot is distortion that precedes a major edge transition expressed as a percentage of Amplitude. The
X cursors show which edge is being measured (edge closest to the trigger reference point).
Figure 11.3
Preshoot
𝑅𝑖𝑠𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑝𝑟𝑒𝑠ℎ𝑜𝑜𝑡 =
𝑙𝑜𝑐𝑎𝑙 𝑀 𝑖𝑛𝑖𝑚𝑢𝑚 − 𝐷 𝑇 𝑜𝑝
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝑋 100
𝐹 𝑎𝑙𝑙𝑖𝑛𝑔 𝑒𝑑𝑔𝑒 𝑝𝑟𝑒𝑠ℎ𝑜𝑜𝑡 =
𝐵𝑎𝑠𝑒 − 𝐷 𝑙𝑜𝑐𝑎𝑙 𝑀 𝑎𝑥𝑖𝑚𝑢𝑚
𝐴𝑚𝑝𝑙𝑖𝑡𝑢𝑑𝑒
𝑋 100
15.
Level@X
: The voltage value between the trigger point and the vertical position of the channel