Rejection seek start offset is now shown on chart as a blue marker when
rejection limits are visible.
Rejection limits (min and max) are now switched when user drags min over max
or vice versa.
TMS Mute
– if set on, TMS Mute is selected.
TMS Decay
– if set on, TMS Decay is selected.
Recalc
– Performs the selected TMS artefact removal operations (TMS Mute & TMS
Decay) on the latest raw (unprocessed) response on demand. The recalculated
response isn’t taken into average, but only shown on screen. This functionality is
meant for adjusting TMS removal parameters at the start of measurement.
Decay fit start (ms)
– See chapter 5.3.6.5.
Coefficients 1 & 2
– See chapter 5.3.6.5.
In curve drawing area of response monitor: right-clicking on top of a single channel in a
monitor opens a settings window for selected channel:
Y-axis max.:
Sets Y-axis maximal level of the channel.
Y-axis min.:
Sets Y-axis minimum level of the channel.
S+
: Increase vertical axis sensitivity of the channel (for quick scale adjustment).
S-
: Decrease vertical axis sensitivity of the channel (for quick scale adjustment).
Set Y-axis range (all):
Y-axis range to all channels according to min
– max
settings.
Set Y-axis range (same signal type):
Y-axis range to all channels with same
signal type according to min
– max settings.
Auto-align
►:
Off
: Auto-align disabled
Bind to single
: Forces the single responses to be shown on the
monitor.