Discover button scans the NeurOne system in order to find out if SyncBox is available and
the number of Main Units in the system and updates that parameter on the UI.
Select the desired Main Unit for Digital OUT definition (the parameter: Selected Main Unit).
Red rectangle indicates the selection. Main Unit number of Master = 0 and unit
number 1…9
stands
for Slave 1… Slave 9.
In case there is only one Main Unit (standalone) without SyncBox in the system, the Master
is the only available choice on the UI and Number of Main Units = 1.
Main Unit specific Settings
Packet frequency (Hz)
The emitted data is delivered at the sampling rate the
measurement is running (i.e. there’s no additional
downsampling performed), at regular intervals. The chosen
delivery rate (i.e. how often UDP datagrams are emitted)
depends on measurement configuration. User can request
delivery rates of 100 Hz, 250 Hz, 500 Hz, 1 kHz, 2 kHz, 3
kHz, 4 kHz and 5 kHz. The higher the delivery rate, the
less latency there’ll be.
Target IP Address
IP Address of the receiving real-time system where the
Digital OUT is transmitting the data packets to.
Make sure that IP address definition in is defined correspondingly
in TCP/IP settings of TARGET PC!
Make sure that Digital OUT cable is connected to the target PC
prior to system start-up.
Target UDP Port
UDP Port of receiving real-time system.
CheckBox:
Send packets MeasurementStart and MeasurementEnd
If checked, MeasurementStart and MeasurementEnd
packets are sent.
MeasurementStart packet is sent by NeurOne when it
starts measuring, before any measurement data (samples
/ triggers) is emitted and also upon reception of Join*
packet.