In this chapter we will look into creating a simple Protocol and using it to conduct a test
measurement. The measurement in question will be a simple auditory evoked potential
P300 measurement. An additional software and hardware for generating the stimulus
audio signals and associated triggers are required to perform the example case. For
example, Presentation software and StimTracker with a response pad.
To begin a new measurement with a brand new measurement Protocol (i.e.
measurement setup) the work begins with the creation of a new Protocol. Open the
Protocol tool by clicking on the Protocol tab on the top left and start a new Protocol by
selecting “New” in the Protocol tool tab. The fields on the right of the screen will activate
and enable you to fill in and select values in the text boxes and drop-down menus.
Name the Protocol as you wish. Set the sampling rate at 5,000 Hz. When you are
finished, proceed by clicking the green “Next” in the bottom row of buttons.
The second window in Protocol creation will prompt you to fill in the sensors you will be
using in the measurement. In this Protocol, we will be using 64 EEG channels. Fill in the
table by selecting the type and name of the sensors. You can jump through the fields by
clicking the tabulator or selecting the wanted field with a mouse. The software will
populate most of the fields for you, but you are free to edit any fields if needed. To add /
delete sensors click Add, Add Many and Delete buttons accordingly. If you already have
a protocol for your cap, you can import the protocol with the Import button. When you
have finished filling in the information, click “Next” to proceed.