Page 182 of 191
800581-2.11 NeurOne System User Manual.doc
Example 2.
Two channels are directed to the digital out and both the sampling rate and
packet delivery frequency are set to be 500 Hz.
Byte
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17
Value
uint8
2
0
0
0
0
0
0
30
0
2
0
1
0
0
0
0
0
Byte
18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34
Value
uint8
0
0
0
30
0
0
0
0
0 234 96 248 231 55 248 232 51
Type
Description
Value
uint8
Value
uint8
Sample packet identifier
2
2
uint8
Main unit indication
0
0
uint8
Not in use
0
0
uint8
Not in use
0
0
uint32
Packet sequency number
0 0 0 30
30
uint16
Number of channels
0 2
2
uint16
Number of sample bundles
0 1
1
uint64
Sample index of the first bundle 0 0 0 0 0 0 0 0 30
24
uint64
Time stamp of the first bundle
0 0 0 0 0 0 234 96 60000
int24
Data sample channel1
248 231 55
-465097
int24
Data sample channel2
248 232 51
-464845