Shock energies / peak voltages at: RV
SVC + housing
Lumax 540 VR-T, VR-T DX, DR-T, HF-T as 40 J device with shock path:
RV
SVC + housing
Shock energies / peak voltages at: RV
SVC
Lumax 540 VR-T, VR-T DX, DR-T, HF-T as 40 J device with shock path: RV
SVC
Behavior in case of EMI
In case of electromagnetic interference, the device switches into the V00 mode when
the interference rate is exceeded if VVI(R) or VDD(R) pacing mode was set previously, or
into the D00 pacing mode if DDI(R) or DDD(R) pacing mode was set previously.
Mechanical Characteristics
Housing
Materials in contact with body tissue
The following materials of the ICD system have contact with body tissue:
Radiopaque marker
The radiopaque marker is on the left, positioned vertically.
Shock energy set
Measurement of shock energy at 50
1 J
E nominal
94%
25%
0.7 ... 1.18 J
E actual
0.81 J
Tolerance range: U max
90 ... 120 V
U max actual
101.01 V
20 J
E nominal
94%
10%
16.9 ... 20.9 J
E actual
17.31 J
Tolerance range: U max
440 ... 480 V
U max actual
465 V
40 J
E nominal
94%
10%
33.8 ... 41.4 J
E actual
35.31 J
Tolerance range: U max
630 ... 670 V
U max actual
656.7 V
Shock energy set
Measurement of shock energy at 50
1 J
E nominal
94%
25%
0.7 ... 1.18 J
E actual
0.83 J
Tolerance range: U max
90 ... 120 V
U max actual
104.2 V
20 J
E nominal
94%
10%
16.9 ... 20.9 J
E actual
17.54 J
Tolerance range: U max
440 ... 480 V
U max actual
466.7 V
40 J
E nominal
94%
10%
33.8 ... 41.4 J
E actual
35.60 J
Tolerance range: U max
630 ... 670 V
U max actual
659.3 V
Device type
W x H x D in mm
Volume in ccm
Mass g
VR-T, VR-T DX, DR-T
66 x 55 x 13
37.2
92
HF-T
66 x 59 x 13
39.8
94
Housing
Header
Sealing plugs, sealing lips
Titanium
Epoxy resin
Silopren
Manufacturing year
Radiopaque marker
nn
BIO SH
Shock energy set
Measurement of shock energy at 50